| In order to investigate the behaviors of estrogens (e.g., E1, E2, E3) in wastewater, it is necessary to determine not only free estrogens but also conjugated estrogens which may be transformed to active estrogens through wastewater treatment. Previously, we reported a procedure for determination both of free estrogens and conjugated estrogens in wastewater by liquid chromatography/tandem mass spectrometry (LC/MS/MS). But the recovery efficiency of conjugated estrogens was particularly low for glucuronated estrogens (less than 30%).In order to overcome, we improved a pretreatment procedure and a subsequent separation by liquid chromatograph. This paper explains an improved analytical method for quantifying conjugated estrogens in wastewater using LC/MS/MS.Summary of the analytical method for wastewater samples are as follows: After addition of an ion-pair reagent and surrogates (E2-S-d4 and E2-G-13C4) to the sample, solid phase extraction (SPE) was performed using Oasis HLB cartridge. Subsequently, the cartridge was eluted with ethyl acetate/methanol solution containing 0.05% ammonia. After solvent removal, the residue was dissolved in acetonitrile/water solution, which was then analyzed by LC/MS/MS.Applying this method, the obtained recovery efficiencies ranged from 80 to 110 % for sulfates conjugates and 60 to 120 % for glucuronides conjugates, respectively.Another objective of this research was to evaluate the efficiency of the two methods. Samples were collected at influent and effluent of five wastewater treatment plants. The following conjugated estrogens were measured: E1-S, E2-S, E3-S, E1-3G, E2-G, E3-G. The improved method was proved to analyze conjugated estrogens in wastewater more efficiency than the existing method except for E1-S. |